: The sensitivities and specificities of three-dimensional time-of-flight MR angiography (3D-TOF MRA) and 3D digital subtraction angiography (3D-DSA) were compared for evaluation of cerebral aneurysms after endosaccular packing with Guglielmi detachable coils (GDCs). Thirty-three patients with 33 aneurysms were included in this prospective study. 3D-TOF MRA and 3D-DSA were performed in the same week on all patients. Maximal intensity projection (MIP) and 3D reconstructed MRA images were compared with 3D-DSA images. The diameters of residual/recurrent aneurysms detected on 3D-DSA were calculated on a workstation. In 3 (9%) of 33 aneurysms, 3D-TOF MRA did not provide reliable information due to significant susceptibility artifacts on MRA. The sensitivity and specificity rates of MRA were 72.7 and 90.9%, respectively, for the diagnosis of residual/recurrent aneurysm. The diameters of residual/recurrent aneurysms that could not be detected by MRA were significantly smaller than those of detected aneurysms (mean 1.1 vs mean 2.3 mm). In one aneurysm of the anterior communicating artery (ACoA), the relationship between the residual aneurysm and the ACoA was more evident on MRA than DSA images. MRA can detect the recurrent/residual lumen of aneurysms treated with GDCs of up to at least 1.8 mm in diameter. 3D-TOF MRA is useful for follow-up of intracranial aneurysms treated with GDCs, and could partly replace DSA.