Background: Current guidelines recommend pulmonary-vein isolation by means of catheter ablation as treatment for drug-refractory paroxysmal atrial fibrillation. Radiofrequency ablation is the most common method, and cryoballoon ablation is the second most frequently used technology.

Methods: We conducted a multicenter, randomized trial to determine whether cryoballoon ablation was noninferior to radiofrequency ablation in symptomatic patients with drug-refractory paroxysmal atrial fibrillation. The primary efficacy end point in a time-to-event analysis was the first documented clinical failure (recurrence of atrial fibrillation, occurrence of atrial flutter or atrial tachycardia, use of antiarrhythmic drugs, or repeat ablation) following a 90-day period after the index ablation. The noninferiority margin was prespecified as a hazard ratio of 1.43. The primary safety end point was a composite of death, cerebrovascular events, or serious treatment-related adverse events.

Results: A total of 762 patients underwent randomization (378 assigned to cryoballoon ablation and 384 assigned to radiofrequency ablation). The mean duration of follow-up was 1.5 years. The primary efficacy end point occurred in 138 patients in the cryoballoon group and in 143 in the radiofrequency group (1-year Kaplan-Meier event rate estimates, 34.6% and 35.9%, respectively; hazard ratio, 0.96; 95% confidence interval [CI], 0.76 to 1.22; P<0.001 for noninferiority). The primary safety end point occurred in 40 patients in the cryoballoon group and in 51 patients in the radiofrequency group (1-year Kaplan-Meier event rate estimates, 10.2% and 12.8%, respectively; hazard ratio, 0.78; 95% CI, 0.52 to 1.18; P=0.24).

Conclusions: In this randomized trial, cryoballoon ablation was noninferior to radiofrequency ablation with respect to efficacy for the treatment of patients with drug-refractory paroxysmal atrial fibrillation, and there was no significant difference between the two methods with regard to overall safety. (Funded by Medtronic; FIRE AND ICE ClinicalTrials.gov number, NCT01490814.)

Copyright (C) 2016 Massachusetts Medical Society. All rights reserved.